CRONO SC PORTABLE PUMPS
for palliative treatment, pump therapy and asthma

The portable CRONO SC and CRONO 50 SC pumps are designed for subcutaneous infusion of palliative care drugs in terminal patients and for subcutaneous infusion of Treprostinil

The pump is indicated in subcutaneous infusions allowing 4 modes of administration:

  1. Continuous
  2. Bolus on request
  3. Clinical bolus (managed by the doctor)
  4. Combined (continuous + bolus on demand + clinical bolus)

The pump is equipped with a keypad lock to prevent incorrect or unauthorised administration.+

The liquid crystal display shows physician and patient information on pump programming, operation and diagnostics.

Crono SC. Infusion pump for palliative treatment, pump therapy and asthma

Canè Crono SC pump

CRONO SC ambulatory pump has been designed for the subcutaneous infusion of drugs for the palliative care of terminal patients and for the subcutaneous infusion of Treprostinil.
The pump is indicated for subcutaneous use and allows for 4 administration methods:

  • 1. Continuous
  • 2. Bolus on request
  • 3. Clinical bolus (managed by the doctor)
  • 4. Combined (continuous + bolus on request + clinical bolus).

The pump is equipped with a settings lock to prevent incorrect or unauthorised administration.

The liquid crystal display shows practical information to the doctor and patient about the settings, operation and diagnostics of the pump.

REQUEST INFO TECHNICAL DATA SHEET

Technical Specifications

  • Pump dimensions: 76 x 49 x 29 mm (3.00 x 1.94 x 1.15 in).
  • Pump dimensions (with protective wings:) 138 x 49 x 29 mm (5.43 x 1.94 x 1.15 in).
  • Pump weight: 115 g (including battery).
  • Weight (with protective wings:) 127 g (4.46 oz.), including battery.
  • Battery: 3 Volt Lithium battery, mod. 123 A (battery life approximately 100 infusions).
  • Single-use reservoirs: Special 10 and 20 ml reservoirs with “Luer-Lock” universal safety fitting.
  • Administration doses: Selectable, from 1 ml to 20 ml in 1 ml increments.
  • Priming function: Only available at the start of the infusion. The maximum volume that can be delivered is 1500 μl in 5 μl increments.
  • Flow rates: Selectable from 5 to 2000 μl/h in 1 μl/h increments:
    It is possible to disable this function by setting 0 μl/h
  • Bolus dose: Selectable, from 10 μl to 2000 μl in 10 μl increments.
    It is possible to disable this function by setting 0 μl.
  • Clinical bolus: Selectable, from 10 μl to 2000 μl in 10 μl increments.
    It is possible to disable this function by setting 0 μl.
  • Minimum interval between bolus doses: Selectable, from 5 min to 24 hours in the following modes:
    - from 5 minutes to 1 hour in 5 minute increments;
    - from 1 hour to 24 hours in 15 minute increments.
    It is possible to disable this function by setting no,Lt.
  • Bolus doses in one hour: From 0 to 12 bolus doses.
    The function may be disabled by setting no,Ln.
  • Flow rate precision: +/-2%.
  • Maximum occlusion pressure: 2,5 bar +/- 1,5 (PL1), 4 bar +/-2 (PL2), 5,5 bar +/- 2 (PL3).
  • Display: Liquid crystal display (dimensions 1.1 x 2.8 cm).
  • Motor: Coreless DC motor. The microcontroller controls the rotation speed using an infrared encoder.
  • Safety circuits: They check that the device is functioning correctly, intervening in the event of any fault with acoustic signals and messages on the display.
  • Anti free-flow system: To avoid a non controlled flow inside the delivery set due to the action of the force of gravity.
  • Supplied standard equipment: Ambulatory infusion pump, infusion pump carry-case, elastic belt, collar strap, fabric case, 2 batteries, battery cover opening tool, 2 user manuals (doctor and patient).
  • Guarantee: Two years for manufacturing defects.

Accessories

Crono SC with protective wings closed. Infusion pump for palliative treatment, pump therapy and asthma
Crono SC with protective wings. Infusion pump for palliative treatment, pump therapy and asthma
Crono SC with stand. Infusion pump for palliative treatment, pump therapy and asthma

Canè Crono 50 SC pump

CRONO 50 SC is a new and innovative PCA pump that is greatly appreciated by both patients and health professionals for its technical characteristics and reduced size.
CRONO 50 SC is an ambulatory infusion pump for controlled administration of drugs in hospitalised patients or those undergoing home treatment.
The pump is indicated for subcutaneous use allowing 4 administration methods:

  • 1. Continuous;
  • 2. Bolus on request (PCA);
  • 3. Clinical bolus (managed by the doctor);
  • 4. Combined (continuous + bolus on request + clinical bolus).

The pump is equipped with a settings lock to prevent incorrect or unauthorised administration. The liquid crystal display shows practical information to the doctor and patient about the settings, operation and diagnostics of the pump.

REQUEST INFO TECHNICAL DATA SHEET
Crono 50 SC. Infusion pump for palliative treatment

Technical Specifications

  • Pump dimensions: 84 x 55 x 42 mm
  • Pump weight: 139 g (including battery).
  • Battery: 3 Volt Lithium battery, mod. 123 A (battery life approximately 100 infusions).
  • Dedicated single-use reservoirs: With a 50 ml capacity and a “Luer-Lock” universal security attachment.
  • Partial volume: Selectable, from 1 to 50 ml in 1 ml increments.
  • Priming function: Only available at the start of the infusion. The maximum volume that can be delivered is 1.5 ml.
  • Flow rates: Selectable from 0.05 to 5.00 ml/h in 0.01 ml/h increments.
    It is possible to disable this function by setting 0 ml/h.
  • Bolus dose: Selectable from 0.10 to 2 ml in 0.02 ml increments.
    It is possible to disable this function by setting 0 ml.
  • Clinical bolus: Selectable from 0.10 to 2 ml in 0.02 ml increments.
    It is possible to disable this function by setting 0 ml.
  • Minimum interval between bolus doses: Can be selected from 5 min to 24 h in the following ways:
    - from 5 min to 1 h in 5 min increments;
    - from 1 h to 24 h in 15 min increments.
    It is possible to disable this function by setting no,Lt.
  • Bolus doses in one hour: From 0 to 12 bolus doses.
    It is possible to disable this function by setting no,Ln.
  • Flow rate precision: +/-3%.
  • Maximum occlusion pressure: 1,8 bar +/-0,8 (PL1) - 2,2 bar +/-1 (PL2) - 3,2 bar +/-1,2 (PL3).
  • Setting memory: The settings are automatically entered in a flash memory which is retained (with the exception of the partial syringe volume) even if the battery is removed from the device.
  • Display: Liquid crystal display (0.43 x 1.1 in, 1.1 x 2.8 cm).
  • Motor: Coreless DC motor. The microcontroller controls the rotation speed using an infrared encoder.
  • Safety circuits: They check that the device is functioning correctly, intervening in the event of any anomaly with acoustic signals and messages on the display.
  • Anti-free-flow safety system: To avoid a non controlled flow inside the delivery set due to the action of the force of gravity.
  • Supplied standard equipment: Ambulatory infusion pump, infusion pump carry-case, elastic belt, collar strap, fabric case, 2 batteries, battery cover opening tool, 2 user manuals (doctor and patient).

Accessories